Keysight E4428C ESG RF Programming Manual page 65

Signal generators
Table of Contents

Advertisement

8. Sets the MXG how the lists will be entered:
Send the following SCPI command:
:LIST:TYPE LIST
An arbitrary list will be used instead of range and step size arguments.
9. Send the MXG a list of frequencies:
Send the following SCPI command:
:LIST:FREQ 100MHz,200MHz,300MHz,400MHz,500MHz
The MXG will put out signals at these frequencies and in this order.
10. Send the MXG a list of powers:
Send the following SCPI command:
:LIST:POW 0dBm,-1dBm,-2dBm,-3dBm,-4dBm
The MXG will put out signals at these powers and in this order.
11. Turn off MXG signal modulation:
Send the following SCPI command:
:OUTP:MOD OFF
12. Turn MXG signal output on:
Send the following SCPI command:
:OUTP ON
13. Configure the MXG's LXI Output LAN Event:
Send the following SCPI commands:
:LXI:EVENt:OUTPut:LAN:DRIVe "LAN0",NORMal
:LXI:EVENt:OUTPut:LAN:SLOPe "LAN0",POS
:LXI:EVENt:OUTPut:LAN:SOURce "LAN0","Waiting For Trigger"
Together, these commands will cause the MXG to send a "LAN0" peer to peer message every time
the "Waiting For Trigger" instrument event changes state.
14. Set the MXG's trigger source:
Send the following SCPI command:
:LIST:TRIG:SOUR LAN
15. Enable the MXG's LXI LAN Event trigger:
Send the following SCPI command:
:TRIG:LXI:LAN:ENABled "LAN1",1
16. Enable the MXG's LXI Output LAN Event:
Send the following SCPI command:
:LXI:EVENt:OUTPut:LAN:ENABled "LAN0",1
Keysight Signal Generators Programming Guide
Using IO Interfaces
Using LAN
57

Advertisement

Table of Contents
loading

Table of Contents