25. Make sure the MXA's LXI LAN triggers are disabled:
Send the following SCPI command:
:TRIG:LXI:LAN:DISable:ALL
26. Set the MXA's trigger source to LXI LAN:
Send the following SCPI command:
:LIST:TRIG:SOUR LAN
27. Configure the MXA's LXI Output LAN Event:
Send the following SCPI commands:
:LXI:EVENt:OUTPut:LAN:DRIVe "LAN1",NORMal
:LXI:EVENt:OUTPut:LAN:SLOPe "LAN1",POS
:LXI:EVENt:OUTPut:LAN:SOURce "LAN1","WaitingForTrigger"
This will cause the MXA to take a measurement at the first frequency in its list. It will also begin
waiting for the "LAN0" peer to peer message before moving on to the next measurement.
28. Enable the MXA's LXI LAN Event trigger:
Send the following SCPI commands:
:TRIG:LXI:LAN:ENABled "LAN0",1
29. Enable the MXA's "OperationComplete" LXI Output LAN Event:
Send the following SCPI commands:
:LXI:EVENt:OUTPut:LAN:ENABled "OperationComplete",1
The MXA will send an "OperationComplete" peer to peer message when the
"OperationComplete" instrument event changes state.
30. Begin the MXG's list sweep:
Send the following SCPI commands:
:INIT:CONT ON
This will cause the MXG to put out a signal at the first frequency in its list. It will also begin
waiting for the "LAN1" peer to peer message before moving on to the next signal.
31. Begin the MXA's list sweep:
Send the following SCPI commands:
:INIT:LIST
This will cause the MXA to take a measurement at the first frequency in its list. It will also begin
waiting for the "LAN0" peer to peer message before moving on to the next measurement.
32. Enable the MXA's "LAN1" LXI Output LAN Event:
Send the following SCPI commands:
:LXI:EVENt:OUTPut:LAN:ENABled "LAN1",1
The "LAN1" peer to peer message is now enabled.