Hide thumbs Also See for 16902A:
Table of Contents

Advertisement

3 Data will appear in the Listing Window upon completion of the run.
4 From the Main Menu choose Markers→New.
a You can accept the default name for the new marker.
b Change the Position field to Value.
c Select the Occurs... button and create the marker setup shown below.
Click
here
to add
event
5 In the Value window, select the Properties... button.
6 In the Value Properties window, select Stop repetitive run when value is not
found.
7 Select OK to close the marker Value Properties window.
8 Select OK to close the marker Value window. The system will search the
display for the occurrences specified.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 3: Testing Logic Analyzer Performance
Test Pod 1 in 300 Mb/s Mode
Click
here
to select
"Or"
47