is -2.9 ns. Note that in this step, you place the blue bars in the narrow
window (not the wide window) that appears to the left of zero in the Eye
Finder display. Then run Eye Finder. The position may be different based
on your test setup. Bring stray channels into alignment if necessary. See
page 69.
18 In the "Setup and Trigger..." window, Sampling tab, Clock Setup area, set
the J clock mode to "Both Edges."
19 In the Eye Finder window, align the blue bars vertically. See page 44.
20 Grab the blue bar for "Label1 (4 Channels)" and move it to the
recommended starting position you noted in the prior step.
21 Run Eye Finder again. Some eyes may close, but the eyes in the sampling
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 3: Testing Logic Analyzer Performance
Test Pod 1 in 600 Mb/s Mode
81

Advertisement