Chapter 3: Testing Logic Analyzer Performance

Test Pods 3 and 4 in 600 Mb/s Mode

Test the complement of the bits (Pod 2, 600 Mb/s mode)

1 Test the complement of the bits on Pod 2. You can use the procedure "Test
the complement of the bits (300 Mb/s mode)" on page 49 as a guideline.
2 Upon completion, the logic analyzer is completely tested.
Test Pods 3 and 4 in 600 Mb/s Mode
Pods 1 and 2 must be used for time tag storage when using Pods 3 and 4 in 600
Mb/s mode.
1 Select the Overview tab.
2 Select My 16950A→Setup→PodAssignment...
3 Reserve Pods 1 and 2 for timetag storage, and assign Pods 3 and 4 to My
16950A. The acquired data will be erased.
4 Re-configure the markers to search for 131071 occurrences instead of
131072 (see "Configure the markers" on page 46).
5 Perform the normal and complement tests for each additional pod on the
logic analyzer, changing the connection to the pod, channel assignments,
thresholds, etc. as appropriate. You must use Clk1 on Pod 1 for all tests in
58