Chapter 3: Testing Logic Analyzer Performance
Test Pod 1 in 300 Mb/s Mode
9 Select OK to close the New Marker window.

Determine PASS/FAIL (2 of 2 tests)

Pass/Fail Point: The Listing window is set up to search for the appropriate
number of A's and 5's in the acquisition. If the logic analyzer does not
detect the correct number of A's and 5's, an error window will appear.
1 Select the Run Repetitive icon
minute. The analyzer will acquire data and the Listing Window will
continuously update.
If the "can't find occurrence" window appears, then the logic analyzer fails
the test.
Check your test setup. If the failure is not the result of a problem with the
test setup, record the failure in the "Minimum Master to Master Clock Time/
Minimum Pulse Width" section of the Performance Test Record.
Be sure that the black ground clip is making good contact with the ground
NOTE:
pin on the test connector.
2 After about one minute, select the Stop button
If the "can't find occurrence" window does not appear, then the analyzer
passes the test. Record "Pass" in the "Minimum Master to Master Clock Time/
Minimum Pulse Width" section of the Performance Test Record.