Configure the 16900-Series Logic Analysis System

1 Exit the logic analysis application (from the main menu, choose File→Exit)
and then restart the application. This puts the logic analysis system into its
initial state.
2 Disable all logic analyzers other than the analyzer under test.
a Select the Overview tab at the bottom of the main window.
b Click on each unused logic analyzer and select disable. Only the logic
analyzer to be tested should remain enabled.
3 Set up the bus and signals.
a From the Logic Analysis System main menu, select Setup→My
16950A→Bus/Signal....
b In the Analyzer Setup window, choose the Threshold button for Pod 1.
The Threshold Settings window will appear.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 3: Testing Logic Analyzer Performance
Testing the Module Using a 16900-Series Mainframe