2 Solder the pin strip to the SMA board mount connector:
a Solder the leads on the left side of the pin strip to the center conductor
of the SMA connector as shown in the diagram below.
b Solder the leads on the right side of the pin strip to the inside of the
SMA connector's frame as shown in the diagram below. Use a small
amount of solder.
3 Attach the second SMA board mount connector:
a Re-heat the solder connection made in the previous step, and attach
the second SMA connector, as shown in the diagram below. Note that
the second SMA connector is upside-down, compared to the first. Add a
little solder to make a good connection.
b Solder the center conductor of the second SMA connector to the center
conductor of the first SMA connector and the leads on the left side of
the pin strip.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 3: Testing Logic Analyzer Performance
Assemble the SMA/Flying Lead Test Connectors
solder