Chapter 3: Testing Logic Analyzer Performance
Test Pod 1 in 300 Mb/s Mode
message is displayed the test passes.
If the red error window "Error- Listing <1> Pattern NOT found..." appears
during the minute or so that it runs repetitively, then the logic analyzer
fails the test. Record the failure in the appropriate place on the
Performance Test Record.
Be sure that the black ground clip is making good contact with the ground
NOTE:
pin on the test connector.
2 After approximately 1 minute click the red stop button in the Listing
window.