page 45.
21 Select OK to close the Eye Finder window.
22 Set the Clk1 mode to "Both Edges."
23 Open the Eye Finder window, and align the blue bars vertically. See page
44.
24 Grab the blue bar for "My Bus 1" and move it to the recommended starting
position you noted in the prior step.
25 Run Eye Finder again. Some eyes may close, but the eyes in the sampling
position you chose on page 54 should remain open.
When you close the Analyzer Setup window a dialog will appear. Answer
erase the data and continue.
26 Perform the procedure "Determine PASS/FAIL (1 of 2 tests)" on page 46.
27 Select the Run Repetitive icon
28 Perform the procedure "Determine PASS/FAIL (2 of 2 tests)" on page 48.

Test the complement of the bits (Pod 1, 600 Mb/s mode)

Now test the logic analyzer using complement data.
1 On the 8133A pulse generator, in the PULSE setup for CHANNEL 2, select
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 3: Testing Logic Analyzer Performance
Test Pod 1 in 600 Mb/s Mode
.
Yes
to
55