tested, then select Individual in the pop-up menu.
g Using the mouse, un-assign all data channels on all pods. Now assign
channels 2, 6, 10, and 14 for the pod being tested. An asterisk (*)
means that a channel is assigned.
h Select OK to close the channel assignment window.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 3: Testing Logic Analyzer Performance
Testing the module using a 16700-series Mainframe
channel
assignment
field
scroll
bar
65