Agilent Technologies 16902A Service Manual page 30

Logic analyzers
Hide thumbs Also See for 16902A:
Table of Contents

Advertisement

3 Set up the oscilloscope.
a Set up the oscilloscope according to the following tables.
Oscilloscope Setup
Setup: Channel 1
On
Scale: 50 mV/div
Offset: 1 V
Coupling: DC
Input: 50 ohm
Setup: Channel 3
Off
Setup: Horizontal
Scale: 500 ps
Position: 525 ps
Reference: Center
Delayed: not selected Edge: Rising Edge
Measure: Markers
Mode: Manual
placement
All else: (n/a)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 3: Testing Logic Analyzer Performance
Setup: Ch. 1 Probe
Setup: Channel 2
Attenuation: 1.00:1
On
Units: Volts
Scale: 50 mV/div
Attenuation Units:
Offset: 1 V
Ratio
External Gain: (n/a)
Coupling: DC
Skew: (Set later. See
Input: 50 ohm
page 33.)
External Offset: (n/a)
Setup: Channel 4
Off
Setup: Trigger
Setup: Acquisition
Mode: Edge
Sampling Mode:
Equiv. Time
Source: Channel 1
Memory Depth:
Automatic
Level: 1.00 V
Averaging: Enabled
# of Averages: 4
Sweep: Auto
Set Up the Test Equipment
Setup: Ch. 2 Probe
Attenuation: 1.00:1
Units: Volts
Attenuation Units:
Ratio
External Gain: (n/a)
Skew: 0.0 seconds
External Offset: (n/a)
Setup: Display
Waveforms: Connect
dots
Persistence: Minimum
Grid: On (and set
intensity)
Backlight Saver: as
preferred
29