3-7 TO CONNECT THE OUTPUT TERMINAL

The Output terminal are as follows.
Terminal block
OUT + : Positive side of the voltage output/current output(±10 V or 4 to 20 mA)
OUT - : Negative side of the voltage output/current output (the same potential as the COM)
The WGA-120A is able to change the output signal (voltage/current) by using the internal jumpers.
For details, see "2. INITIAL SETTINGS".
Output signal (voltage): The output load should be 2k ohm or more.
Output signal (current): The output load should be 500 ohm or less.
Take countermeasures against noise when in outdoors or when in noisy areas due to motors, etc.