KYOWA WGA-120A Instruction Manual page 47

Carrier type instrumentation amplifier
Table of Contents

Advertisement

[Example 1] Actual load calibration 1
To measure weight by using single load cell (rated capacity: 1kN)...
First, with the no load applied, turn the Sensitivity adjuster all the way to the right to execute the balance
adjustment.
Then, with the weight (1kN) applied to the load cell, adjust the output signal to be 1.00 V or 10.00 V for
reading the weight directly.
Suppose the WGA-120A outputs 1.00 V when applying the weight (1kN).
When the WGA-120A outputs 0.80V, the applied weight is 0.8kN.
-42-