KYOWA WGA-120A Instruction Manual page 54

Carrier type instrumentation amplifier
Table of Contents

Advertisement

Output signal to be output (V) × CAL (mV/V) × calibration constant (kN/mV/V)
Output signal (V) =
5.00 (V) × 0. 712 (mV/V) × 0.5005 (kN/mV/V)
=
= 3.564 (V)
NOTE
The above procedures indicate how to adjust the ZERO by using the output signal (voltage).
The WGA-120A is also able to adjust the ZERO by using the output signal (current). The
output signals (current: 4 to 20 mA) move in tandem with the output signals (voltage: 0 to
+10). (For details, see "2. INITIAL SETTINGS.")
Actual load (kN)
0.5 (kN)
-49-