[Example 2] Actual load calibration 2
Load cell
Rated capacity: 1kN
Assume that 5 V output signal is required when 1kN is loaded. When applying the weight (0.8kN), the
output signal (voltage, V) is as follows.
0.8×5
Output =
= 4 (V)
1
First, with the no load applied, turn the Sensitivity adjuster all the way to the right to execute the balance
adjustment.
Then, adjust the output signal to be 4.00 V when applying the weight (0.8kN).
Be sure to use the weight near the max. weighing capacity. Or, the precision of measurements will be
decreased.