3-6 TO CONNECT THE AUTO BALANCE CONTROL INPUT TERMINAL

The Auto balance control input terminal (AZ and COM) on the terminal block is designed to execute the
auto balance adjustment.
Use a contact point (switch or relay).
Terminals for executing the auto balance adjustment are as follows.
Terminal block
AZ:
Auto balance control input
COM: Common
For how to use the auto balance adjustment, see
"4-5 AUTO BALANCE ADJUSTMENT."
NOTE
The function is not able to use on the manual balance model.