[Example 2] How to use the CAL
Assume that a load cell is used to obtain 5.00 V of output signal when 0.5kN is loaded. Calculate the
output signal (voltage) after adding the CAL.
<Conditions>
Load cell
CAL of the WGA-120A
Rated capacity:
Bridge resistance:
Calibration constant:
*Calibration constant: Unique value described on the test data sheet
0.25mV/V (350-ohm bridge), that is, 0.712mV/V (1000-ohm bridge, see
the table 4-1)
-48-
1kN
1000 ohm
0.0005005kN/1μV/V
(= 0.5005kN/1mV/V)