3.19 Conditioning
■■■■■■■■
48
3 Starting equilibration
Start the equilibration in MagIC Net: Workplace ▶ Run ▶ Equili-
■
bration ▶ Start HW.
Visually inspect whether all capillaries and their connections from
■
the high-pressure pump to the detector are leak-tight. If eluent is
leaking out anywhere, tighten the corresponding pressure screw
further, or loosen the pressure screw, check the end of the capil-
lary and shorten it using the capillary cutter if necessary and
retighten the pressure screw.
4 Conditioning the system
Continue rinsing the system with eluent until the desired stability
level for the baseline has been attained .
The instrument is now ready for measuring samples.
■■■■■■■■■■■■■■■■■■■■■■