3.11 Metrohm Suppressor Module (MSM)
3.11
Metrohm Suppressor Module (MSM)
■■■■■■■■
28
Exchanging the sample loop
NOTE
Observe a dead-volume-free installation of the sample loop (see "Creat-
ing dead-volume-free capillary connections", page 15).
1 Removing the existing sample loop
Loosen the pressure screws (6.2744.010) at Port 3 and Port 6.
■
Remove the sample loop.
■
2 Installing a new sample loop
Fasten one end of the sample loop to Port 3 using a PEEK pressure
■
screw (6.2744.010).
Use the second PEEK pressure screw (6.2744.010) to fasten the
■
other end of the sample loop to Port 6.
The suppressor drive of the 930 Compact IC Flex can hold various rotors.
The large rotors, such as the MSM‑HC Rotor A (6.2842.000) and the
MSM-HC Rotor C (6.2842.200) can be inserted directly.
The small rotors, such as the MSM Rotor A (6.2832.000) and the MSM-LC
Rotor A (6.2844.000), must first be fitted into the adapter (6.2842.020),
which can then be inserted into the suppressor housing.
A connecting piece (6.2835.010) is used for all rotors for connecting the
Metrohm Suppressor Module (MSM) to the IC system.
■■■■■■■■■■■■■■■■■■■■■■