930 Compact IC Flex Oven/ChS (2.930.2200)
2 Mounting the tubing adapter
Install the parts of the tubing adapter for aspiration filter
(6.2744.210) accessory set:
Start by pushing the tubing weighting onto the loose end of the
eluent aspiration tubing.
Then push the clamping screw onto the loose end of the eluent
aspiration tubing.
Lastly, push the filter holder onto the loose end of the eluent aspi-
ration tubing and screw it onto the tubing nipple.
The end of the tubing should extend approximately 1 cm.
3 Pre-rinsing the aspiration filter
NOTE
Always wear gloves when handling the aspiration filter.
In order to avoid air bubbles after the installation of the aspiration fil-
ter, we recommend pre-rinsing the aspiration filter with water or elu-
ent.
For pre-rinsing, you will need the holder for the eluent aspiration fil-
ter (6.2744.360), a syringe and the purge needle (6.2816.040).
This procedure can also be found in the video "Inserting a new aspi-
ration filter" at the following link:
4 Mounting aspiration filter
NOTE
Always wear gloves when handling the aspiration filter.
Place the loose end of the eluent aspiration tubing into the aspira-
tion filter.
The end of the tubing should reach approximately to the center of
the aspiration filter.
Tighten the aspiration filter to the filter holder.