Metrohm 930 Compact IC Flex Oven/ChS Manual page 49

Table of Contents

Advertisement

■■■■■■■■■■■■■■■■■■■■■■
930 Compact IC Flex Oven/ChS (2.930.2200)
3 Starting equilibration
Start the equilibration in MagIC Net: Workplace ▶ Run ▶ Equili-
bration ▶ Start HW.
4 Deaerating the high-pressure pump
Push the end of the purge needle (6.2816.040) over the end of
the purge capillary on the purge valve.
Insert the syringe (6.2816.020) in the Luer connector of the purge
needle.
Open the purge valve using the rotary knob (approx. ½ turn).
Switch on the high-pressure pump in MagIC Net.
Use the syringe to aspirate eluent until there are no more air bub-
bles in the eluent aspiration tubing.
Switch off the high-pressure pump in MagIC Net.
Seal the purge valve using the rotary knob.
Remove the syringe from the purge needle.
Pull the purge needle out of the purge capillary.
5 Rinsing the instrument without columns
Rinse the instrument (without columns) with eluent for 10
minutes.
3 Installation
■■■■■■■■
41

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 930 Compact IC Flex Oven/ChS and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents