Thermo Scientific 43iQTL Instruction Manual page 99

Trace level sulfur dioxide analyzer
Table of Contents

Advertisement

Optical Bench Settings
Thermo Scientific
The Optical Bench Settings screen allows the user to calibrate the optical
bench. The calibration in this screen should be performed if the part
associated with the optical bench has been changed/repaired or if the span
coefficient is outside the limit of 0.5 to 2.0.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue To Detector Gain Settings
Screen
The Detector Gain Settings screen contains the following information:
Across:
G
Description: Define the actions the user can do.
G
Detector Gain: User can manually set or automatically calibrate the
G
detector gain.
Operation
Settings
3-73
43iQTL Instruction Manual