Pump Replacement - Thermo Scientific 43iQTL Instruction Manual

Trace level sulfur dioxide analyzer
Table of Contents

Advertisement

Replacement
Thermo Scientific
Pump
Use the following procedure to replace the pump.
Equipment required:
Phillips drive, #1 and #2
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2. Unplug pump cable from Step POL board J7.
3. Twist opposite direction to unlock tube clamps.
Note Push in tube clamp to lock.
4. Disconnect tubing from pump.
5. Unfasten two captive hardware.
6. Slide pump left until keyway meets opening.
2–1).
L
43iQTL Instruction Manual
Servicing

Pump Replacement

7-19