Permeation Oven Board Replacement; Figure 7-53. Replacing The Permeation Oven Board - Thermo Scientific 43iQTL Instruction Manual

Trace level sulfur dioxide analyzer
Table of Contents

Advertisement

Permeation Oven
Replacement
Thermo Scientific
Use the following procedure to replace the permeation oven board.
Board
Equipment required:
Phillips drive, #2
CAUTION Allow oven to cool down prior to servicing.
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2. Unplug all electrical J1, J7, J8, J9 as shown.
3. Using #2 Phillips drive, unfasten four 6-32 pan head screws.
Figure 7–53. Replacing the Permeation Oven Board
4. Replace permeation board and assemble in reverse order.
2–1).
Servicing

Permeation Oven Board Replacement

L
43iQTL Instruction Manual
7-49