Table 5-10. "Get Status" Command Return Data In A "Dual-Smartshutter" Configuration - Sutter Instrument LAMBDA 10-B Operation Manual

Optical filter changer and smartshutter® control system
Table of Contents

Advertisement

56

Table 5-10. "Get Status" command return data in a "Dual-SmartShutter" configuration.

Byte
Label
Offset
0
Command echo
1
Shutter A
open/closed state
2
Shutter B
open/closed state
3
Shutter A mode
(SmartShutter
only)
4
(5)
5 or 6
Shutter B mode
(SmartShutter
only)
6 or 7
(7 or
8)
7, 8,
Return data
or 9
terminator
* NOTE: The block of data returned by the Get Status command will vary in length depending on whether neither, either, or
both SmartShutters (A and/or B) is/are in Neutral Density mode. All modes are followed by an additional byte indicating which
device to which the mode pertains (1 = Shutter A and 2 = Shutter B. When Neutral Density mode is selected, a third byte exists
that contains the number of microsteps (1 – 144) used to indicate the Neutral Density level. The following table outlines the
characteristics of the status block of data according to the mode of each of the two SmartShutters.
Byte Value
Dec. Hex.
Binary
204 CC 11001100
170 AA 10101010
171 AB 10101011
172 AC 10101100
186 BA 10111010
188 BC 10111100
219 DB
11011011
220 DC 11011100
221 DD 10111011
222 DE 10111100
1
01
00000001
1
01
00000001
-
144
90
10010000
219 DB
11011011
220 DC 11011100
221 DD 10111011
222 DE 10111100
2
02
00000002
1
01
00000001
-
144
90
10010000
13
0D
00001101
LAMBDA 10-B OPERATION MANUAL – REV. 1.20Q (20230227)
Description
The Status command byte code
echoed back.
Open
Open is conditional
Closed
Open
Closed
SmartShutter A not connected:
Indicates either that no
SmartShutter is connected, or a
Vincent shutter is connected.
Fast Mode
Soft Mode
Neutral Density Mode
Device designator: Shutter A
A third byte contains the number
of microsteps (1 – 144).
NOTE: The number of neutral
density microsteps is present at
this byte offset only if Shutter A
Mode is set to Neutral Density
(Offset 3 = 222 (0xDE)).
SmartShutter B not connected:
Indicates either that no
SmartShutter is connected, or a
Vincent shutter is connected.
Fast Mode
Soft Mode
Neutral Density Mode
Device designator: Shutter B
A third byte contains the number
of microsteps (1 – 144).
NOTE: The number of neutral
density microsteps is present at
this byte offset only if Shutter B
Mode is set to Neutral Density
(Offset 5 or 6 = 222 (0xDE)).
ASCII CR (carriage return).
Bit Values
Dec./–
Binary
Hex

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LAMBDA 10-B and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents