57
5.4.2 All Motors Power On
This command is used to instruct the controller to switch ON electrical power to the motors
of all connected filter wheels and shutters.
5.4.3 All Motors Power Off
This command is used to instruct the controller to switch OFF electrical power to the motors
of all connected filter wheels and shutters.
5.4.4 Transfer to On Line
This command is the equivalent to pressing the ON LINE key on the LAMBDA 10-B's
keypad. If the controller is in local mode (i.e., it is off line), this command will cause the
controller to go on line (the opposite of local mode), enabling it to respond to other
commands from the remote computer.
5.4.5 Transfer to Local
This command is the opposite of the Transfer to On Line command. It causes the controller
to go into local mode and is the equivalent to pressing the LOCAL key on the controller's
keypad. Once in local mode, the controller will not respond to any commands sent to it from
the remote computer except the Transfer to On Line command.
5.4.6 Reset
Causes the LAMBDA 10-B controller to reset, changing all settings back to their defaults.
After the reset has completed, status information similar to that returned by the Status
command is returned to the remote computer.
5.4.7 Get Controller Type and Configuration
This command is used to obtain information about the controller as to its model and
configuration. The following table shows the type of information returned when issuing this
command.
LAMBDA 10-B OPERATION MANUAL – REV. 1.20Q (20230227)