10.1 Technical Specifications

All specifications herein mentioned apply to the DSO1000E series oscilloscopes. Before checking
an oscilloscope to see if it complies with these specifications, make sure it meets the following
conditions:
The oscilloscope must have been operating continuously for twenty minutes under the
specified operating temperature.
The Do Self Cal operation must be performed through the Utility menu if the operating
temperature changes by more than 5℃.
The oscilloscope must be within the factory calibration interval.
All specifications are guaranteed unless noted 'typical'.
Oscilloscope Specifications
Horizontal
Sample Rate Range
Waveform Interpolation
Record Length
TIME/DIV Range
Sample Rate and
Delay Time Accuracy
Delta Time Measurement
Accuracy
(Full Bandwidth)
Position Range
DSO1000E Series HandHeld Oscilloscope User Manual
Specifications
1GS/s
(sin x)/x
Maximum 2M samples per single-channel; maximum 1M
samples per dual-channel (4K, 40K, 512K, 1M optional)
DSO1072E
DSO1102E
4ns/div to 2Ks/div, in a 2, 4, 8
sequence
± 50ppm over any ≥1ms time interval
Single-shot, Normal mode
± (1 sample interval +100ppm × reading + 0.6ns)
>16 averages
± (1 sample interval + 100ppm × reading + 0.4ns)
Sample interval = s/div ÷ 200
DSO1072E DSO1102E
4ns/div to 8ns/div
20ns/div to 80μs/div
200μs/div to 2Ks/div
DSO1152E DSO1202E
2ns/div to 10ns/div
Specifications
DSO1152E
DSO1202E
2ns/div to 2Ks/div, in a 2, 4,
8 sequence
(-8div × s/div) to 20ms
(-8div × s/div) to 40ms
(-8div × s/div) to 2Ks
(-4div × s/div) to 20ms
76