Cursor - Hantek DSO1000E Series User Manual

Handheld oscilloscope
Table of Contents

Advertisement

waveform display can be scaled or positioned by vertical and horizontal controls.
Equivalent Acquisition: Just repeat the Normal acquisition. Use this mode to take a specific
observation on repeatedly displayed periodic signals. You can get a resolution of 40ps, i.e.
25GSa/s sample rate, which is much higher than that obtained in real-time acquisition.
The acquisition principle is as follows.
Input repeated signals
First Acquisition
Second Acquisition
Third Acquisition
Fourth Acquisition
As shown above, acquire input signals (cycle repeatable) for more than once at a slow sample rate,
arrange the sample points by the time they appear, then recover waveforms.

5.5.6 CURSOR

The Cursor Menu.
Options
Settings
Off
Type
Voltage
Time
CH1
CH2
Source
MATH
REFA
REFB
Select
S
Cursor
E
Display the
difference
Delta
(delta)
between the
cursors.
Moving Cursors: Press the key F3 to select a cursor and move it with the Arrow key. Cursors can
be moved only when the Cursor Menu is displayed.
DSO1000E Series HandHeld Oscilloscope User Manual
Select a measurement cursor and display it.
Voltage measures amplitude while Time measures time and
frequency.
Select a waveform to take the cursor measurement.
Use the readouts to show the measurement.
S indicates Cursor 1. E indicates Cursor 2.
A selected cursor is highlighted, which can be moved freely. Both
cursors can be selected and moved at the same time. The box
behind the cursor displays the location of the cursor.
Display the measurement in the box under this option.
Comments
Basic Operation
37