Otto Bock 1C11 Terion K2 Instructions For Use Manual page 13

Hide thumbs Also See for 1C11 Terion K2:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 10
5 Preparing the product for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
NOTICE
Altering the prosthetic foot or footshell
Premature wear resulting from damage to the product
► Do not alter the prosthetic foot or footshell.
INFORMATION
There is a plastic protector on the product's pyramid. It protects the con­
necting section from scratches during the alignment and fitting of the pros­
thesis.
► Remove the protector before the patient leaves the workshop/fitting
area.
5.1 Applying/removing the footshell
INFORMATION
► Pull the protective sock over the prosthetic foot to prevent noises in the
footshell.
► CAUTION! Always use the prosthetic foot with the footshell.
Apply or remove the footshell as described in the footshell instructions
for use.
5.2 Bench Alignment
TT bench alignment
Required materials: 662M4 goniometer, 743S12 heel height measuring
apparatus, 743A80 50:50 gauge, alignment apparatus (e.g. 743L200
L.A.S.A.R. Assembly or 743A200 PROS.A. Assembly)
Assemble and align the prosthetic components in the alignment apparatus
according to the following specifications:
Heel height: Effective heel height (shoe heel height - sole thick­
   
ness of forefoot) + 5 mm
Bench alignment process
13

Advertisement

Table of Contents
loading

Table of Contents