6
•
Operation
6.3
General Guidelines for Operation
Page 72
Consider the information and guidelines for operation to optimize
detector performance.
•
Keep the device doors closed during operation to avoid exposure of
the module's front to mechanical shock or thermal changes.
•
Ensure that the operating conditions are suitable. This includes:
Stable environmental conditions, such as a stable temperature
¨
No air drafts
¨
No vibrations or mechanical shocks caused by external sources
¨
No EMC-related sources of strong interference other than, for
¨
example, certified laboratory equipment.
Stable backpressure and correct waste line setup
¨
TIP During measurements, avoid touching the module's front (close the
doors) or the electrical connectors as this may impair the measurement
signal.
Refractive Index Detector (VC-D60)
Operating Manual