Thermo Scientific VC-D60 Operating Manual page 102

Refractive index detector
Table of Contents

Advertisement

7
Maintenance and Service
Page 102
2. Push all tubing and capillaries, which are present in the tubing chase
of the Vanquish system modules, into the tubing chase. Otherwise,
you will not be able to remove the slide-in module properly from the
enclosure in the next step.
3. Grasp the slide-in module by the leak tray and pull the module out
of the enclosure by approximately 10 cm.
Figure 25: Pulling out the slide-in module
NOTICE
The slide-in module can fall down when pulling it out of the enclosure
too far.
Pull out the slide-in module just far enough so that you can grasp it on
both sides from below.
4. Remove the slide-in module from the enclosure. The following steps
require a team effort:
a) Take the slide-in module on both sides from below.
b) Pull the slide-in module from the rails towards the front.
c) Place the slide-in module on a clean and stable surface.
5. If the module is transported over a longer distance or if it is shipped:
Fasten the locking screws (see
(} page 98)).
Tightening the Locking Screws
Refractive Index Detector (VC-D60)
Operating Manual