Thermo Scientific VC-D60 Operating Manual page 62

Refractive index detector
Table of Contents

Advertisement

5
Installation
Connecting the inlet capillary from the UV/VIS detector
Page 62
2
Figure 15: Connecting the inlet capillary
NOTICE
Be aware of the backpressure limit of the flow cell in the Vanquish UV/
VIS detector (refer to the Operating Manual for the UV/VIS detector) in
the flow path before the refractive index detector.
Avoid connecting any additional components in the flow path between
the two detectors and connect the capillary from the UV/VIS detector
directly to the refractive index detector flow cell inlet.
3
Figure 16: Connecting the inlet capillary from the UV/VIS detector
(example)
1. Connect the inlet capillary from the column compartment to the
flow cell inlet of the UV/VIS detector. Refer to the instructions in the
Operating Manual for the UV/VIS detector.
2. Route the connecting capillary from the flow cell outlet of the UV/
VIS detector upward to the refractive index detector.
1
4
2
Refractive Index Detector (VC-D60)
Operating Manual
1