Thermo Scientific VC-D60 Operating Manual page 96

Refractive index detector
Table of Contents

Advertisement

7
Maintenance and Service
Page 96
1. Use the screwdriver to remove the fuse holder.
2. Replace the two fuses with new fuses of the specified type and
current rating. Always replace both fuses.
3. Reinstall the fuse holder.
4. Reconnect the power cord to the power source and to the device.
5. Turn on the device with the main power switch.
Refractive Index Detector (VC-D60)
Operating Manual