Thermo Scientific VC-D60 Operating Manual page 64

Refractive index detector
Table of Contents

Advertisement

5
Installation
Follow these steps
Page 64
TIP Store the flow cell plugs, for example, in the drawer of the system
base, to have them easily available when storing or shipping the
detector.
TIP Connect the refractive index detector as the last module in the flow
path. This is due to the outlet pressure limit of the detector (0.05 mPa or
less).
1. Connect the waste line to the flow cell outlet (OUT).
1
Figure 18: Connecting the detector waste line
2. Position the tubing bracket in the middle of the enclosure height to
establish a slightly elevated routing of the waste line. Route the
waste line behind the tubing bracket as shown in the figure.
3. Guide the waste line downward in the tubing guides of the system
modules to the Vanquish system base.
4. On the system base, route the detector waste line through the
dedicated detector waste outlet and connect the detector waste line
to the waste container (refer to the Vanquish System Operating
Manual).
TIP The waste line should go straight to the system base and to waste.
Make sure that the line is positioned straight in the tubing guides.
Refractive Index Detector (VC-D60)
Operating Manual
2