Settings For Calibration - Xylem WTW Turb 750 IR Operating Manual

Lab turbidimeter
Table of Contents

Advertisement

®
Turb
750 IR/T
Settings
ba77233e05
06/2023

4.9.1 Settings for calibration

The Calibration menu item is displayed if the authorization exists
(see section 4.2).
The settings are in the menu Configuration / Turbidity / Calibration.
To switch to the Configuration menu, press the <MENU
Menu item
Calibr. record
Calibr. type
Standard
Calibr. interval
Possible
Explanation
setting
-
Display the calibration record of
the last calibration.
The selected calibration type is
3-P StdCAL
used to calculate the current
PharmaCAL
measured value.
FreeCAL
The active calibration type is indi-
QuickCAL
cated in the measured value dis-
play, and stored together with a
measured value.
When you start calibration with
the <CAL
started with the selected calibra-
tion type.
1... 5
Number
Number of calibration standards
for the Calibr. type FreeCAL
Standard 1
Here the nominal values are en-
tered for all calibration standards
...
Standard 5
1... 999 d
Calibration interval for turbidity
measurement (in days).
If the calibration interval has ex-
pired, the meter reminds you to
calibrate before each measure-
ment.
Operation
(7)
> key.
(4)
> key, it is always
35