Reset; Resetting The System Settings - Xylem WTW Turb 750 IR Operating Manual

Lab turbidimeter
Table of Contents

Advertisement

®
Turb
750 IR/T
Example:
output format CSV
ba77233e05
06/2023
The CSV output format delivers datasets separated by ";".
The data are output in the following order:
1
2
1
2
15.01.20;15:06:49;1;26.5;NTU;VALID;[AQA][3-P StdCAL];15.01.20;08:67:47
15.01.20;16:06:49;1;16.5;NTU;VALID;[AQA][3-P StdCAL];15.01.20;08:67:47
1 Date of storing
2 Time of storing
3 Selected ID
4 Measured value or
Upper/lower measuring range limit
(only with measured value status, OFL/UFL)
5 Unit of the measured value
6 Measured value status
* VALID: Measured value valid
* INVALID: Measured value invalid
* UFL: Measured value below the lower measuring range limit
* OFL: Measured value above the upper measuring range limit
7 AQA status
* [AQA]: a valid AQA check is available
*
: no valid AQA check is available
8 Calibration status: Calibration type that was selected for the
measurement
9 Calibration status: Date of calibration
10Calibration status: Time of calibration
If the connected external printer does not print, please check
whether the same baud rate is set at the Turb
printer.

4.13 Reset

You can reset (initialize) all system and measurement settings.

4.13.1 Resetting the system settings

With the System / Reset function, all resettable settings are reset.
 Settings for Turbidity (see section 4.13.2)
 System settings
3
4
5
6
3
4
5
6
7
8
9
7
8
9
®
750 IR/T and
Operation
10
10
55

Advertisement