Conditioning, calibration, measurement
General informa-
tion
Before measuring
16
Conditioning, calibration, measurement
When operating the electrode ensure that
the filling opening (1) for the electrolyte is open
no air bubbles are in the electrolyte
the depth of immersion is within the optimum range:
Minimum depth of im-
mersion
Maximum depth of im-
mersion
1
Condition the electrode in standard solution prior to using.
To achieve a good responding behavior, the concentration of the
standard solution should be similar to the concentration expected of
the test sample.
2
For calibrating and measuring, the plug of the refilling opening of the
electrode has to be open.
The membrane and junction have to be immersed in the test sample.
3
Remove any air bubbles in the electrolyte by slightly knocking
against the shaft.
4
Connect the electrode to the meter
(see operating manual of your meter).
5
Calibrate (see operating manual of your meter).
1 Filling opening
2 Filling level of the electrolyte
1
1
3 Junction
2
3
The junction (3) must be covered
Approx. 1 cm below the fluid level (2) of the
electrolyte
Na 800
ba77237e02