Operation
Settings
30
Calibration
Standards
type
(FNU/NTU)
FreeCAL
2 ... 5

4.7.1 Settings for calibration

The settings are in the menu Konfiguration / Trübung / Kalibrierung.
To switch to the Konfiguration menu, press the <MENU> key.
Menu item
Kalibrierprotokoll
Kal.-typ
Standard
(only with Kal.-typ
FreeCAL)
Explanation
Guided flexible calibration with 2 to 5
freely selectable calibration standards.
The nominal values for the calibration
standards are entered manually.
The order of the entered nominal values
corresponds to the order of the calibra-
tion standards during the calibration
process.
Possible
Explanation
setting
-
Display the calibration record of
the last calibration.
3-P StdCAL
The selected calibration type is
used to calculate the current
FreeCAL
measured value.
QuickCAL
The active calibration type is indi-
cated in the measured value dis-
play, and stored together with a
measured value.
When you start calibration with
the <CAL/ZERO> key, it is al-
ways started with the selected
calibration type.
Anzahl
1 ... 5
Number of calibration standards
for the Kal.-typ FreeCAL
Standard 1
Here the nominal values are en-
tered for all calibration standards
...
Standard 5