What to do if...

Display, LoBat
Meter does not
react to keystroke
Error message
i AQA not
activated!
Error message,
Error
0, 8, 16, 16384
You want to know
which software
version is in the
meter
60
6
What to do if...
6.1

General information

Cause
– Batteries almost empty
Cause
– Software error
– Operating condition undefined
or EMC load unallowed
Cause
– The AQA function was not
activated.
Cause
– Instrument error
Cause
– E.g., a question by the service
department
Remedy
– Insert new batteries
Remedy
– Processor reset:
Press the <START/ENTER
(8)
<PRT
> keys simultaneously.
Remedy
– Activate the AQA-check (see
section 4.10.2 S
ETTINGS FOR
QUALITY ASSURANCE
Remedy
– Repeat measurement
– Meter defective,
contact the service department
Remedy
– Switch on the meter.
– Open the menu Configuration/
System / Info. The instrument data
are displayed.