4.9.3 Carrying out a calibration

For quick and easy measuring we recommend that you use marked cells with
the calibration standards (see section 4.5.1 P
If your cells are not marked you can determine the suitable position of the in-
dividual cells in the cell shaft while measuring.
Perform the following preparatory activities when you want to calibrate:
1
Select the calibration type (menu Configuration / Turbidity / Calibr.
type.).
For QuickCAL calibration with the <QuickCAL
setting of the QuickCAL calibration type and limitation of
the measuring range are done automatically.
2
Clean the cell (see section 5.2.2).
3
Insert the cell (see section 4.4).
4
Press the <CAL
The guided calibration with the selected calibration type starts.
Follow the instructions on the display.