Care and Maintenance
Adapter to 1 mm Output Port (without ruggedized adapter)
It is recommended that you always connect a ruggedized adapter to the U9361M output port.
However, if you want to connect an adapter that uses the inner threads, or cable directly to a
test port connector, use the 4 lb-in end of the dual torque wrench (silver end of the dual torque
wrench). Do not use the red end of the dual torque wrench which is only used when connecting
the stronger outside test port connector threads to the ruggedized adapter.
Never apply in excess of 4 lb-in of torque to the U9361M output port inner threads.
Figure 10
Wrenches Positioned on Correct Flats
Torque wrench
8710-2819
4 lb-in
torque wrench
U9361M
Output Port
(1.0 mm)
To connect a cable or device to the 1 mm adapter, use a 6 mm open end wrench to hold the
test port adapter stationary, and torque the cable or device to 4 inch pounds (0.45 N-m).
A common 1mm cable connection nut is 6 mm.
For the U9361F, with the 2.4 mm output connection, use the 5/16 inch torque wrench to torque
the cable or device to 8 inch-pounds (0.90 N-m).
32
Groove marks the correct place
to hold wrench
1.0 mm
connector
U9361C/F/G/M Receiver Calibrator Quick Start Guide
6 mm open-end
wrench
adapter
(not supplied
with kit)