Keysight Technologies U9361G Quick Start Manual page 23

Receiver calibrator
Hide thumbs Also See for U9361G:
Table of Contents

Advertisement

8.
To apply the correction, check the "Apply" column in the row. The table should appear as:
The "Applied" column should show "Yes", indicating that the selected calibration is being
applied. For more detailed information, press the RCal Status button to bring up the status
page:
The status for this Cal Group and Row is Calibration Status "Calibrated" and Apply Status
"Applied", indicating that the calibration was successful and the cal has been applied to the
measurement state. Press Close to return to the swept analyzer display.
This completes a basic magnitude calibration from 1 GHz to 3 GHz using the RCal module.
U9361C/F/G/M Receiver Calibrator Quick Start Guide
Example: RCal Calibration Using a UXA Signal Analyzer
23