Keysight Technologies U9361G Quick Start Manual page 27

Receiver calibrator
Hide thumbs Also See for U9361G:
Table of Contents

Advertisement

Connection Procedure
Familiarize yourself with the proper use of the double-ended torque wrench shown in Figure 5
below. To prevent expensive repairs and repair downtime, take a few minutes to study the
information in the following pages.
Figure 5
Dual Torque wrench, 14 mm open end, 4 lb-in (0.45 Nm)
and 10 lb-in (1.13 Nm), Part Number 8710-2819
Test Port Connector
Figure 6 shows the torque specification for the two connection threads on the U9361M output
port connector.
Two wrenches must be used to avoid stress on cables.
Figure 6
Torque Specification for Output Port Threads
U9361C/F/G/M Receiver Calibrator Quick Start Guide
FOR KEYSIGHT RUGGED 1mm
MATING CONNECTORS ONLY
FRONT
BACK
BACK
10 lb-in torque for
ruggedized adapters
(outer threads)
4 lb-in torque
(inner threads)
Care and Maintenance
27

Advertisement

Table of Contents
loading

This manual is also suitable for:

U9361fU9361cU9361m

Table of Contents