Example: RCal Calibration Using a UXA Signal Analyzer
Procedure
1.
On the analyzer select Mode/Meas > Spectrum Analyzer Mode > Swept SA Measurement.
Then, press Mode Preset.
2.
Select: Input/Output > Calibration > Configuration. This should bring up the RCal
Configuration table:
3.
Press Select Calibrator, and replace None with RCal Module 1.
The model number and serial number should now be displayed below the Calibrator option.
These numbers should match the numbers physically on the device.
If RCal Module 1 does not show up as an option, disconnect and reconnect
the USB cable.
4.
Press Use Current Meas to bring up a row in the table.
5.
In the row, select:
Start Freq > 1 GHz
Stop Freq> 3 GHz
20
U9361C/F/G/M Receiver Calibrator Quick Start Guide