Getting Started

1. Start the HP VEE on your PC.
2. Load the adjustment program le into the HP VEE as follows:
a. Pull down the File menu from the HP VEE window and select Open .
b. Select the le C:\E4915_16\ADJUST.VEE .
c. Click
NNNNNNNN
OK .
Note
You may be asked GPIB addresses of the test equipment during the program
loading. Enter the address for each equipment. Enter 0 as the address
for the equipment which are not used for the adjustments/performance
tests. The GPIB address for each test equipment can be changed in the
adjustment/performance test program.
NNNNNNNNNNNNNNNNN
3. Click
Start button on the Adjustment/Performance Test Program screen.
4. In the \Setup" window, modify the settings if required.
5. Click
NNNNNNNN
OK .
6. Modify the calibration factor of the HP 8482A in the window.
7. Click
NNNNNNNN
OK .
8. Select the model number and click
9. Enter the serial number of the E4915A/E4916A exactly. For example,
10. Mark installed options of the E4915A/E4916A. The installed options can be con rmed on
the rear pannel of the E4915A/E4916A, or when you turn the E4915A/E4916A ON, you
can see the installed options before the E4915A/E4916A starts measurement.
Note
The installation status of the E4916A Option 001 Add Impedance Probe cannot
be con rmed on the rear pannel or the power-on message. You have to ask the
user if the Impedance Probe is used with the E4916A.
NNNNNNNN
11. Click
OK .
NNNNNNNN
OK .
JP1KD00123
Adjustments and Performance Tests 2-7