Agilent Technologies E4915A Service Manual page 103

Crystal impedance / lcr meter
Table of Contents

Advertisement

-313
Calibration memory lost
The nonvolatile calibration data has lost.
-350
Queue overflow
A speci c code entered into the queue in lieu of the code that caused the error. This code
indicates that there is no room in the queue and an error occurred but was not recorded.
-400
Query error
This is the generic query error that the E4915A/E4916A cannot detect more speci c errors.
This code indicates only that a error as de ned in IEEE 488.2, 11.5.1.1.7 and 6.3 has occurred.
-410
Query INTERRPUTED
A condition causing an interrupted error occurred (see IEEE 488.1, 6.3.2.3); for example, a
query followed by DAB or GET before a response was completely sent.
-420
Query UNTERMINATED
A condition causing an unterminated query error occurred (see IEEE 488.2, 6.3.2); for example,
the E4915A/E4916A was addressed to talk and an incomplete program message was received.
-430
Query DEADLOCKED
A condition causing an deadlocked query error occurred (see IEEE 488.2, 6.3.1.7); for example,
both input bu er and output bu er are full and the E4915A/E4916A cannot continue.
-440
Query UNTERMINATED after indefinite response
A query was received in the same program message after an query requesting an inde nite
response was executed (see IEEE 488.2, 6.5.7.5.7).
Messages-7