Isolation Test (Isolation); Required Equipment; Procedure - Agilent Technologies E4915A Service Manual

Crystal impedance / lcr meter
Table of Contents

Advertisement

7. Isolation Test (ISOLATION)

Required Equipment

50 Termination BNC Type

Procedure

1. Connect the equipment as shown in Figure 2-9.
2. Select \ISOLATION" and click
3. Following the program instructions, complete the procedure.
2-16 Adjustments and Performance Tests
11593A,
E4915A: 2ea, E4916A: 4ea
Figure 2-9. Isolation Test Setup
NNNNNNNN
OK in the list box of the program display.