2.6.3 Leaving Program Mode

To leave the program mode, press the key, "prog" (11). Then, inside the display "time", appears the
word "programm".
Set the display to "programm--" (M10) with the control field (1).
For all number marked text, please refer to Figure 20.

2.7 Starting and Stopping the Centrifuge

2.7.1 Starting the Centrifuge

Start the centrifuge with either the "start" key (9), or the "quick" key (8). With the "start" key (9),
stored runs or runs with manually pre-selected parameters can be started. During the centrifugation
the LED light is flashing slowly with red color. When the respective pre-selected running time has
ended, the centrifuge will stop automatically.
With the "quick" key (8), start runs, which will last a few seconds, can be initiated. By pressing the
"quick" key (8), the centrifuge accelerates up to the pre-selected revolution. In the display "time" (A-
3), the passed running time is indicated from the moment the "quick" key (8) is pressed. By releasing
the "quick" key (8), the centrifuge stops and the running time is indicated, until the lid is opened.
For all number marked text, please refer to Figure 21.
Figure 21
20
Z287A_V1.19_
eng
© Hermle Labortechnik GmbH