OPERATING
Figure 15

2.5 Radius Correction

If adapters or reducers are being used, it could change the centrifugal radius of the respective rotor.
In that case, the radius can be corrected manually.
Please proceed as follows:
First close the centrifuge lid and press afterwards the key "time" (6) and the key "prog" (11) at the
same time and hold down.
In the display, "time" (A-3), appears the word "radius" (M9). With the control field (1), pre-select the
respective radius correction, (see Table 3, APPENDIX P. IV), in steps of 0.1cm.
As soon as the radius correction is set, the word "radius" (M9) appears. This text remains visible
until the radius correction is set back to 0.
For all number marked text, please refer to Figure 16.
Figure 16

2.6 Program

2.6.1 Storage of Programs

The program stores up to 99 runs, with all relevant parameters, including the used rotors. Any free
program number is available and can be retrieved.
Put the rotor into the centrifuge. By pressing the key, "prog" (11), in the display "time" (A-3) appears
the word "program", (see Figure 17). With the control field (1), choose the desired program number.
18
Z287A_V1.19_
eng
© Hermle Labortechnik GmbH